Model
4102
CLINICAL USE
Lower operating costs are
possible, using the long life Boehringer Inspiratory Force Meters.
Recorded measurements of inspiratory force provide objective indices
of a patients strength; therefore, the measurements can be of
great value, not only in assessing the recovery from neuromuscular
blockade, but also in following the progression of neuromuscular
weakness, as in myasthenia gravis. The recording Inspiratory Force
Meter can be used to check endotracheal tube cuff pressure with an
adapter which is included in the kit.
The recording Inspiratory
Force Meter is designed with a number of outstanding features for the user:
FEATURES
-
A memory pointer captures
the highest inspiratory or expiratory pressure.
-
A removable manifold is
provided with 15 mm. I.D. and 22 mm. O.D. connections. The manifold
has a safety port, which must be willfully occluded by the operator.
The manifold may be removed for steam, gas or chemical sterilization.
Extra manifolds available, Model #9019 (see below). To avoid
contamination, use Isolation Kit, Model # 9016 (see below)
-
A Luer adapter, length of
tubing and three-way stopcock are also included, to measure other
pressures such as endotracheal tube cuff pressure.
ADVANTAGES
-
EtO Sterilizable: 80°F
for 8 ½ hours cold cycle
-
Silicone damped, lightweight
memory pointer simplifies reading
-
Safety ported for failsafe operation
-
May be used in Bain circuit
SPECIFICATIONS
-
RANGE: Model 4102--± 60 cm. H2O, 4103--± 150 cm. H2O
-
ACCURACY: ± 3% full scale
-
FITTING: 15 mm.-22 mm.
-
STERILIZATION: 100 EtO
cycles minimum
-
DIAMETER: 2.75" LENGTH:
5.35" WIDTH: 1.625" WEIGHT: 10 oz.
ACCESSORIES
-
GAUGE PROTECTOR
-
Black neoprene Gauge
Protector shields your Inspiratory Force Meter, should you drop it.
-
½" shock absorbing
lip protects the dial face.
-
Nylon lanyard allows you to
carry meter around your neck.
-
ISOLATION KIT
-
Allows separation of meter
from patient. Useful in septic patient to prevent contamination of meter.
-
Use with Inspiratory Force,
Meter, any model.
-
18" Isolation hose.
-
Sterilizable by steam, gas
or chemical (Cidex) means.
-
EXTRA MANIFOLD
-
To avoid cross
contamination, you may want each patient in an ICU to have his own Manifold.
-
Nylonvirtually unbreakable.
-
Safety ported.
-
15-22 mm. adaptable.
-
Sterilizable by steam, gas
or chemical (Cidex) means.
-
Model #9019C incorporates a